Gas Analyzer, O2 Precision – Gold Edition (#17518)

$6,395.00

O2 Precision Gas Analyzer – Gold Edition (#17518)

This precision analyzer has a range from 0 to 100%, other ranges are available on request. It is typically used to determine the O2 concentration in respired gases or gas mixtures from collection bags, or other sources.

  • Oxygen concentration is displayed on a digital meter
  • Built-in gas pump draws the gas through the analyzer
  • A flow meter indicates the gas sample flow
  • Precision gas mixtures are available for calibrating the analyzer
  • 1-year warranty
SKU: 17518 Category:

Description

# 17518 O2 ANALYZER, GOLD EDITION

A precision paramagnetic oxygen gas analyzer

This precision analyzer has a range from 0 to 100%. It is typically used to determine the O2 concentration in respired gases or gas mixtures from collection bags, or other sources.

The O2 concentration is displayed on a digital meter. With the end-tidal (EtO2) option the lowest value during a breathing cycle may be displayed. Corresponding analog voltages are available on rear panel BNC connectors.

A gas pump is built in to draw the gas through the analyzer.

A flow meter indicates the gas sample flow. Gas flow is internally adjustable. Adjustment may affect response time. A gas return port on the rear panel is standard.

Precision gas mixtures are available for calibrating the analyzer.

 

Meter Display:
    0 – 100.00%
Repeatability:
    • +/- 0.01%
Linearity:
    • +/- 0.1% O
Zero Stability:
    (At Constant Temperature & Constant Flow)  +/- 0.002% O2/hour
Temperature Coefficient:
    • (Within a range of +/-10 degree C from calibration temperature)
    • Zero +/- 0.03 +/- 0 O2/ degree c
  • Typical Span +/- 0.1% FS/degree C Typical
Response Time T90, N2 to 100% O2:
    2.0 sec at 400ml/min flow
Type:
    Paramagnetic
Digital Panel Meter Readout:
    4-1/2 digit LED
Signal Out X1:
    • 0 – 5 VDC = 0 to 100% O
  • 0 – 10 VDC on request, no charge
Signal Out X4:
    • 0 – 5 VDC = 0 to 25% O
Output Noise:
    Less than 2mv peak to peak
Operating Voltage:
    110/220 VAC, 50/60Hz
Size:
    5-3/4″ (146mm) Height x 8-5/8″ (229mm) Width x 15″ (380mm) Depth
Sample Gas Inlet:
      • Rear Panel

    (Front panel on request, no charge)

Drying Column:
    Nafion semi-permeable membrane brings sampled gas to ambient humidity. No chemical dryer required.

Fast Response Option

Model 17518C-FR uses an optional fuel cell sensor with a response time of about 150 ms. Some of above specifications do not apply. 1 year warranty details in FAQ